Omega Vehicle Security OMB-CHARTSCAN 1400 User Manual

Page of 273
Appendix A
API Commands
ChartScan User’s Manual
A-7
 
Command
 
Syntax
 
Description
 
D#
 
Set Relay Make Time
 
D#make
 
Specify how long to wait after closing a channel relay before
reading valid data, where 
make
 is the number of 520.833
microsecond intervals.
 
 
 
 
 
E
 
End Calibration Mode
 
E
 
Terminate Calibration Mode
 
 
 
 
 
E?
 
Error Status Query
 
E?
 
Returns present error condition of the unit with one of the following
error codes:
 
E000
 - No error has occurred.
 
E001
 - Invalid device dependent command (IDDC).
 
E002
 - Invalid device dependent command option (IDDCO).
 
E004
 - Channel configuration error.
 
E008
 - Calibration error.
 
E016
 - Trigger overrun.
 
E032
 - Open TC or Range Error.
 
E128
 - Command conflict error.
 
 
 
 
 
F
 
Set Data Format
 
Fengr,
format
 
Configure the input and output formats.
 
engr
 converts the raw data to the engineering units that you
prefer to work with.  Valid options are: 
0
 - degrees Celsius
(default), 
1
 - degrees Fahrenheit, 
2
 - degrees Rankine, 
3
 -
degrees Kelvin, and 
4
 - Volts.
 
format
 places the readings in a format that you prefer to work
with.  Valid options are: 
0
 - Engineering Units (default), 
1
 -
Binary (Low Byte/High Byte), 
2
 - Binary (High Byte/Low Byte),
and 
3
 - Counts (ASCII).
 
 
 
F?
 
Query the data format.
 
 
 
 
 
F#
 
Set Burst Mode
Frequency
 
F#freq
 
Select burst mode sampling frequency where 
freq
 is a real
number and where 
38.5
 < 
freq
 < 
20000.0
 Hertz (default).
 
 
 
 
 
G
 
Calibrate Channel Gain
 
Gchan,
type
 
Calibrate the correction gain constants for given option card where
chan
 is the channel number such that
 
1
 < 
chan
 < 
max
,
 
where max = 128 for ChartScan/1400.
 
The chassis is 
chan = 0
.
 
type
 is the channel type.
 
 
 
 
 
H
 
Calibrate Channel
Offset
 
Hchan
 
Calibrate the correction offset constants for given option card
where 
chan
 is the channel number such that
 
1
 < 
chan
 < 
max
, where max = 128 for ChartScan/1400.
 
The chassis is 
chan = 0
.
 
 
 
 
 
I
 
Set Scan Interval
 
Inorm, acq
 
Set the two configurable scan intervals of the unit where 
norm
and 
acq
 are the normal and acquisition scan intervals which
have the form: 
hh:mm:ss.t
 where 
hh
 is hours, 
mm
 is
minutes, 
ss
 is seconds, and t is tenths of a second.
 
 
 
I?
 
Query scan interval selection.
 
 
 
 
 
I#
 
Digital Input Stamping
 
I#state
 
Specify whether or not digital input stamping will be enabled.
state
 may have either of the following values: The default of
0
, which disables digital input stamping, or 
1
 to enable digital
stamping.
 
 
 
 
 
J
 
Calibrate Cold Junction
Offset
 
Jchan,
type,
temp
 
Calibrate the cold-junction offsets for cold-junction compensation
of the thermocouple signals where 
chan
 is the channel number
such that 
1
 < 
chan
 < 
max
,
 
where max = 128 for ChartScan/1400.
 
type
 is the thermocouple channel type.
 
temp
 is the reference temperature currently being applied to the
selected channel, with the form 
nnn.n
 in degrees º C.
 
 
 
 
 
K
 
Enter Calibration Mode
 
Kkey
 
Enter Calibration Mode if the keyword (or password) is correct,
where 
key
 is the keyword in the form of a 5-digit number
nnnnn
.  Therefore the password can be no longer than 5
characters.
 
 
 
K?
 
Query 
Kkey
 as defined above.