Configuration Aspects for Programmers
Appendix B
B-16
ChartScan User’s Manual
With Alarm Stamping:
PRINT#1, “OUTPUT07;A#1X”
‘ Enable alarm stamping
.
‘ Configure and start an acquisition
PRINT#1, “OUTPUT07;R1X”
‘ Request the next scan in the buffer
PRINT#1, “ENTER07”
‘ Retrieve it
INPUT A$
‘
....+0234.20-0019.40+0001.40+0023.60 005 128 032 066
Digital Input Stamping (I#)
Digital Input Stamping (
I#
) allows you to see whether a digital input was “active” or “inactive” at the time of
the scan. This is accomplished by a digital input “ON” (1) or “OFF” (0) code appended to the scan in real time.
Digital Input Stamping can be enabled with the
Digital Input Stamping can be enabled with the
I#1
command and disabled with the
I#0
command. Note that
there are 8 digital inputs.
Note:
Digital Input Stamping is not valid in high-speed, single channel mode (burst mode).
Examples of
I#
command follow.
Without Digital Input Stamping (default):
PRINT#1, “OUTPUT07;I#0X”
‘ Disable digital input stamping
.
‘ Configure and start an acquisition
PRINT#1, “OUTPUT07;R1X”
‘ Request the next scan in the buffer
PRINT#1, “ENTER07”
‘ Retrieve it
INPUT A$
‘
....+0234.20-0019.40+0001.40+0023.60