Omega Vehicle Security OMB-CHARTSCAN 1400 User Manual

Page of 273
Appendix C
Registers, Data Formats, & Queries
ChartScan User’s Manual
C-9
 
Configuring Power-Up State
 ChartScan has an internally stored factory default configuration that can be recalled at any time.  You can
program the ChartScan to power-up in a “user-defined” state, or if desired, can configure the unit to power-up
with the default configuration.
 
 During normal operations the factory defaults may be recalled by issuing the Restore Factory Defaults (*F)
command.  When this command is processed, ChartScan loads the factory default configuration and operates
accordingly.  This command, however, may not be issued while an acquisition is currently configured.
 
 ChartScan is usually configured to power-up with the last known configuration (from when the unit was
powered-down).  Last-configuration power-ups will also occur when a Power-On Reset (*R ) command is used.
You can configure the unit such that it will always power-up with the factory default configuration by issuing the
Power-up Settings (*S1) command. Note that issuing an *S0 (power-up under last-known configuration)
command causes the unit to power up under the last known configuration.
 
 Note:
Configuring ChartScan to power-up under the factory defaults or recalling the factory default
settings will not affect pre-existing calibration factors of the ChartScan or its slave units.
 
 The following table lists the execution steps and effects when the ChartScan is configured to power-up under the
factory default configuration.
 
 
 
Step
 
Command
 
Effect
 
1
 
Q7,0,0,0,0
 
Set general response Term.  to CR-LF/EOI.  All others to None
 
2
 
V0
 
Set the User Term.  to ASCII code 00 (Hex) NUL
 
3
 
F0,0
 
Set the data format to Engr.  Units — Degrees C
 
4
 
M0
 
Set SRQ (SRE) Mask to 000
 
5
 
N0
 
Set Event Status Enable (ESE) to 000
 
6
 
*C
 
Un-configure all channels
 
7
 
Y0,0,0
 
Set acquisition counts to all 0
 
8
 
I00:00:00.1,00:
00:00.1
 
Set both timebase intervals to 0.1 seconds
 
9
 
L1, 0,0
 
Un-configure level
 
10
 
T0,0,0,0
 
Un-configure acquisition triggering
 
11
 
M#0
 
Set measuring mode to line cycle integration/high speed multi-channel
mode
 
12
 
F#20000.0
 
Set burst mode frequency to 20kHz
 
13
 
W#32
 
Set sampling weight to 32