Omega Vehicle Security OMB-CHARTSCAN 1400 User Manual

Page of 273
ChartScan User’s Manual 
ChartView Software Reference    4-39
Burst Mode, Operational Issues
              Acquisition Setup Dialog Box, High-Speed, Single-Channel Mode
The Burst Mode, also referred to as High-speed, single-channel mode, is selected with a radio-button located on
the lower left of the Acquisition Setup Dialog Box.  The following bulleted issues apply to Burst Mode operations.
•  When in the High-speed, single-channel (burst mode), all ChartView display functions are intentionally
disabled.  Data acquisition and storage to disk are the only services that remain functional.  When you
select the High-speed, single channel mode, a Warning message box appears, as indicated in the
following figure.
This Warning Appears Prior to Enabling the Burst Mode
•  The Burst Mode channel will be the only channel enabled in the instrument.  This is regardless of the
ON/OFF status indicated on the Channel and Alarm Setup spreadsheet (located on the Setup Channel
Types, Alarms and Acquisition Parameters
 Window).