Specifications
Dual Input Oscilloscope
4
57
Horizontal
Scope Modes .................................. Normal, Single, Roll
Ranges
Normal:
equivalent sampling ................... 20 ns to 500 ns/div
real time sampling .............................. 1 
µ
s to 5 s/div
Single (real time) ..................................... 1 
µ
s to 5 s/div
Roll  (real time) ......................................... 1s to 60 s/div
Sampling Rate (for both channels simultaneously)
Equivalent sampling (repetitive signals) ........................
up to 1.25 GS/s
Real time sampling:
µ
s to 5 ms/div........................................... 25 MS/s
10 ms to 60 s/div ........................................... 5 MS/s
Time Base Accuracy
Equivalent sampling ..................
±
(0.4% +0.04 time/div)
Real time sampling ....................
±
(0.1% +0.04 time/div)
Glitch Detection ..................
40 ns @  20 ns to 5 ms/div
200 ns @ 10 ms to 60 s/div
Glitch detection is always active.
Horizontal Move........................................... 10 divisions
Trigger point can be positioned anywhere across the
screen.
Trigger
Screen Update...............................Free Run, On Trigger
Source.............................................................. A, B, EXT
EXTernal via optically isolated trigger probe ITP120
(optional accessory)
Sensitivity A and B
@ DC to 5 MHz.............................0.5 divisions or 5 mV
@ 25 MHz .................................................. 1.5 divisions
@ 40 MHz ..................................................... 4 divisions
Slope ................................................... Positive, Negative
Video on A .......................... interlaced video signals only
Modes ............................................... Lines, Line Select
Standards...........................PAL, NTSC, PAL+, SECAM
Polarity .............................................. Positive, Negative
Sensitivity..........................................0.6 divisions sync.