Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Ofs Trnsfer
IS-95 only. This field transfers the time offset value from the time offset 
measurement (displayed in 
TmOfs
) to the 
Time Offset
 field 
automatically. Before selecting 
Ofs Trnsfer
, you must make at least 
one of the following code domain measurements:
• Power
• Timing
• Phase 
Operating Considerations
When using this command programmatically, be sure a value for 
Tm 
Ofs
 has been obtained from a power, timing, or phase measurement. 
(One method of ensuring the measurement has been completed would 
be to send an GPIB command to query the results of the 
Tm Ofs
 field 
before sending the transfer command). 
Screens on which this field is present
COD DOM (Code Domain Analyzer): FP setup controls menu
Output Port
This field selects the output port for the signal from the Test Set’s RF 
and CDMA generators.
RF Out
 selects the RF IN/OUT connector.
Dupl
 selects the DUPLEX OUT connector.
Screens on which this field is present
CDMA GENERATOR
Phase/div
This field sets the display resolution for the code domain phase 
measurement: 1, 2, 5, 10, 20, or 50 milliradians (mrad). This field is 
displayed on the CODE DOM screen’s 
Marker
 menu when the 
Measurement
 field is set to 
Phase
Screens on which this field is present
CODE DOM (Code Domain Analyzer, IS-95 only): marker controls 
menu