63
Operating Overview
Calibrating the Adjacent Channel Power (ACP) Measurement
Calibrating the Adjacent Channel Power 
(ACP) Measurement 
To calibrate an ACP measurement using the internal CDMA generator:
Step 12. Remove power from the RF IN/OUT (or ANT IN) port.
Step 13. Go to the CDMA ANALYZER screen and choose the 
ACP
 measurement.
Step 14. Set the 
Tune Freq
 field (or 
RF Channel
 field) to the center frequency of 
your CDMA signal. 
Step 15. Select the 
ACP Cal
 field and wait until the calibration is complete.