Mintek Pacemaker LV-T User Manual

Page of 172
Stratos CRT-Ps 
Implantable Cardiac Resynchronization Therapy Pacemakers 
 
Stratos 
X-Ray identification 
 
Radiopaque Identification 
A radiopaque identification code is visible on standard x-ray, and 
identifies the pulse generator: 
Stratos LV/LV-T 
 
SV
 
 
 
C
AUTION
 
Lead / CRT-P Compatibility – Because of the numerous 
available 3.2-mm configurations (e.g., the IS-1 and VS-1 
standards), lead/ CRT-P compatibility should be confirmed 
with the CRT-P and/or lead manufacturer prior to the 
implantation of the system. 
IS-1, wherever stated in this manual, refers to the 
international standard, whereby leads and generators from 
different manufacturers are assured a basic fit. [Reference 
ISO 5841-3:1992(E)]. 
 