Barco 4MP User Manual

Page of 76
Coronis Fusion 4MP/6MP
43
Display function selection
T
O
 
SELECT
 
A
 
DISPLAY
 
FUNCTION
 (DF):
1
Open the main menu. 
2
Open the Display Function menu.
3
Select Display Function.
4
Select the desired DF.
5
Return to the Display Function menu.
You can select the following DFs:
DICOM
Select a DICOM display function for most 
medical viewing applications.
The DICOM function results in more visible 
grayscales in the images.
Dynamic Gamma 2.2
Dynamic Gamma 1.8
These are gamma functions that are shifted 
to take into account the non-zero 
luminance of an LCD panel when driven 
with a “black” signal.
They are especially useful in CT applications 
to improve the perception of low 
Hounsfield values.
Native
If you select Native, the native panel 
behavior will not be corrected.
Test
This is identical to the Native DF.
Gamma 2.2
Gamma 1.8
Select one of these display functions in 
case the display is to replace a CRT display 
with a gamma of 1.8 or 2.2 respectively.
User
This is a user-definable DF.