Anritsu 682XXB User Manual

Page of 514
Rise/Fall
Time and
Overshoot
The following procedure lets you measure the rise
time, fall time, and overshoot parameters of the
pulse modulation waveform.
1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing
SYSTEM
,
then
Reset
. Upon reset, the CW Menu is dis-
played.
b. Press
Edit F1
to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
d. Press
Edit L1
to open the current power level
parameter for editing.
e. Set L1 to the maximum leveled output power
level for the instrument being tested (refer to
Table 3-2, page 3-6).
f. Press
MODULATION
, then
Pulse
. The Inter-
nal Pulse Status menu is displayed.
g. At the Internal Pulse Status menu, press
Edit Width
and set pulse width W1 to 5.0
ms.
h. Press
Edit Period
and set the PRI to 0.01 ms.
i. Press
On/Off
to turn pulse modulation on.
2. On the Oscilloscope, turn off CH.4. Observe the
output of the Pulse Detector on the Oscilloscope.
Refer to Figure 3-11 to interpret the detector
waveform. Adjust the Oscilloscope controls to
measure the following waveform parameters.
a. Rise/Fall Time (10 to 90%): <10 ns
b. Pulse Width: 5
ms ±10 ns (5 ns, typical)
c. PRI: 0.010 ms
±10 ns (5 ns, typical)
d. Overshoot: <10%
3. Record the results on the Test Record.
3-40
682XXB/683XXB MM
PERFORMANCE
PULSE MODULATION TESTS:
VERIFICATION
RISE TIME, FALL TIME, OVERSHOOT, AND LEVEL