Anritsu 682XXB User Manual

Page of 514
6-12
REMOVING AND
REPLACING THE
A21-1 PCB
This paragraph provides instructions for removing and replacing the
A21-1 BNC/AUX I/O Connector PCB, which is located on the rear
panel assembly (see Figure 6-4).
Preliminary
Disconnect the power cord from the unit. Remove
the chassis covers as described in paragraph 6-2.
Remove the rear panel assembly as described in
paragraph 6-10.
WARNING
When power is removed from the instrument, always allow
five minutes for the capacitors on the A21 Line Filter/Recti-
fier PCB to discharge before removing the A21-1 PCB.
Procedure
Remove and replace the A21-1 PCB as follows:
Step 1
Using a Phillips screwdriver, remove the
four screws that fasten the A21 shield to
the A21 PCB.
Step 2
Remove the shield and set it aside.
Step 3
Disconnect the ribbon cable connector
from the A21-1 PCB.
Step 4
Using a
ANRITSU
P/N T1451 tool, remove
the dress nuts from the 12 rear panel
BNC connectors.
Step 5
Carefully remove the A21-1 PCB from the
rear panel assembly.
Step 6
To replace the PCB, reverse the removal
process.
682XXB/683XXB MM
6-17
REMOVAL AND
REPLACEMENT PROCEDURES
A21-1 PCB