Test
Procedure
(2 to 10 GHz)
The following procedure lets you measure the 2 to
10 GHz (2.2 to 10 GHz for 68XX5Bs) RF output har-
monic levels to verify that they meet specifications.
1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing
SYSTEM
,
then
Reset
. Upon reset the CW Menu is dis-
played.
b. Press
Edit L1
to open the current power level
parameter for editing.
c. Set L1 to the lesser of +10 dBm or the maxi-
mum leveled power level for the instrument
being tested (refer to Table 3-2, page 3-6).
d. Press
Edit F1
to open the current frequency
parameter for editing.
e. Set F1 to the frequency indicated on the Test
Record.
2. Set up the Spectrum Analyzer as follows:
a. Span: 5 kHz/div
b. CF: Set to the 682XXB/683XXB frequency
value.
c. RBW: 1 kHz
d. Video Filter Wide: On
3. Adjust the Spectrum Analyzer Peaking control
for maximum signal level, then adjust the Refer-
ence Level Control to place the signal at the top
of the screen graticule.
4. Change the Spectrum Analyzer CF to each of the
harmonic frequencies listed on the Test Record
and record the signal levels. Refer to Table 3-4 for
the specified harmonic signal level limits.
5. Repeat steps 1 through 4 for each of the 682XXB/
683XXB CW carrier and harmonic frequencies
listed on the Test Record. Record the harmonic
signal levels on the Test Record.
682XXB/683XXB MM
3-19
PERFORMANCE
HARMONIC TEST: RF OUTPUT
VERIFICATION
SIGNALS FROM 2 TO 20 GHz
Harmonic and Harmonic Related:
500 MHz to
£
2.2 GHz (68XX5B):
<–50 dBc
10 MHz to
£
50 MHz:
<–30 dBc
>50 MHz to
£
2 GHz:
<–40 dBc
>2 GHz (2.2 GHz for 68XX5B)
to
£
20 GHz:
<–60 dBc
>20 GHz to
£
40 GHz:
<–40 dBc
Harmonic and Harmonic Related (Models having a
high-end frequency of >40 GHz and units with Op-
tion 15B at maximum specified leveled output
power):
500 Mhz to
£
2.2 GHz (68XX5B):
<–50 dBc
10 MHz to
£
50 MHz:
<–30 dBc
>50 MHz to
£
2 GHz:
<–40 dBc
>2 GHz (2.2 GHz for 68XX5B)
to
£
20 GHz:
<–50 dBc
>20 GHz to
£
40 GHz:
<–40 dBc
50 GHz units: >40 GHz to
£
50 GHz:
<–40 dBc
60 GHz units: >40 GHz to
£
60 GHz:
<–30 dBc
65 GHz units: >40 GHz to
£
45 GHz:
<–25 dBc
>45 GHz to
£
65 GHz:
<–30 dBc
Non-Harmonics:
500 MHz to
£
2.2 GHz (68XX5B):
<–50 dBc
10 MHz to
£
2 GHz:
<–40 dBc
>2 GHz (2.2 GHz for 68XX5B)
to
£
65 GHz:
<–60 dBc
Table 3-4.
Spurious Signals Specifications