9. On the Spectrum Analyzer:
a. Deselect NOISE LVL.
b. Set Frequency Span to 300 kHz.
c. Adjust the Marker for a 100 kHz offset.
d. Select NOISE LVL.
10. Measure the phase noise level 100 kHz offset
from the carrier frequency. Record the level on
the Test Record.
the Test Record.
11. Repeat steps 1 through 10 for all frequencies
listed on the Test Record.
682XXB/683XXB MM
3-25
PERFORMANCE
SINGLE SIDEBAND
VERIFICATION
PHASE NOISE TEST