9. On the Spectrum Analyzer:
a. Deselect NOISE LVL.
b. Set Frequency Span to 300 kHz.
c. Adjust the Marker for a 100 kHz offset.
d. Select NOISE LVL.
10. Measure the phase noise level 100 kHz offset
from the carrier frequency. Record the level on
the Test Record.
11. Repeat steps 1 through 10 for all frequencies
listed on the Test Record.
682XXB/683XXB MM
3-25
PERFORMANCE
SINGLE SIDEBAND
VERIFICATION
PHASE NOISE TEST