Fluke 1523 User Manual

Page of 36
1523, 1524 Thermometer Readout 
Specifications
28
Temperature, RTD Ranges,  
Table 11 
and Accuracies (ITS-90)
Accuracy ± 
°C 4 Wire Probe 
± 0.011 at –100 °C
± 0.015 at 0 °C
± 0.019 at 100 °C
± 0.023 at 200 °C
± 0.031 at 400 °C
± 0.039 at 600 °C
Resolution:  
0.001°C (0.001°F)
Temperature, Thermistor
Table 12 
Accuracy ± 
°C
± 0.002  at  0 °C
± 0.003 at  25 °C
± 0.006 at  50 °C
± 0.014 at  75 °C
± 0.030 at 100 °C
Resolution:  
0.001 °C (0.001 °F)
Based on a 10 k
Ω (at 25 °C) thermister 
with a beta value of 4000 
Ω. See 