   17 CHECK AND ADJUSTMENT
17-5
17.3.1Checking /Adjusting the Plate Level
Adjustment is required if the axis of the plate level is not perpendicular to the vertical axis.
Check
1) Place the plate level parallel to a line running through the centers of two leveling screws, say, A and 
B.  Use these two leveling screws only and place the bubble in the center of the plate level.  
2) Rotate the instrument 180° or 200g around the vertical axis and check bubble movement of the 
plate level.  If the bubble has been displaced, then proceed with the following adjustment.
Adjustment
1) Adjust the level adjustment capstan screw, with the accessory adjusting pin and return the bubble 
towards the center of the plate level.  Correct only one-half of the displacement by this method.
2) Correct the remaining amount of the bubble displacement with the leveling screws.
3) Rotate the instrument 180° or 200g around the vertical axis once more and check bubble 
movement.  If the bubble is still displaced, then repeat the adjustment.
17.3.2Checking /Adjusting the Circular Level
Adjustment is required if the axis of the circular level is also not perpendicular to the vertical axis.
Check
1) Carefully level the instrument with the plate level only.  If the bubble of the circular level is centered 
properly, adjustment is not required.  Otherwise, proceed with the following adjustment.
Adjustment
1) Shift the bubble to the center of the circular level, by adjusting three capstan adjustment screws on 
the bottom surface of the circular level, with the accessory adjusting pin.
Leveling screw 
Plate level
Leveling screw B
Level adjustment 
capstan screw
Half amount of displacement 
Total amount of displacement 
Capstan adjustment
screws
Bottom of the base
Circular level