3M 965AMS User Manual

Page of 204
24
Voltage Termination
1. The 
Voltage Termination 
option allows you to select 
the input impedance of the 
965AMS digital voltmeter 
(in supported countries 
only). The input impedance 
of the internal 965AMS 
voltmeter is normally 
1Mohm. However, some 
legacy systems use voltage 
measurement systems 
with input impedances of 
100Kohms. This option 
is provided to maintain 
measurement compatibility 
with those systems.
 
If 100Kohm termination is 
selected, the 965AMS tester 
will display ‘100K’ on the 
voltage measurement.
 
Use this menu to change 
the Voltage Termination. 
Use the up and down arrow 
keys to highlight Voltage 
Termination. Press 
Change
 
to continue or press 
 
to quit without making 
changes.
 