Service Procedures  
173
Appendix D
Distortion
Verification
This test checks the Harmonic Distortion at selected frequencies and 
harmonics. This test requires the use of a spectrum analyzer with dynamic 
range, frequency range, and resolution bandwidth adequate for the 
measurement.
Select each function generator output in the table below. Use a spectrum 
analyzer connected to the function generator output. Set the fundamental 
frequency reference to 0 dB and measure the 2nd through 5th harmonic 
frequencies relative to this reference. This test is a 50 
Ω output termination 
test.
Table D-10. Distortion Verification
Agilent E1441A
Measurement
harmonic
Function
 Out
Term
a
Ampl
Freq
Fundamental
2nd 3rd  4th  5th 
Amt 
below
reference
O
Sine wave
50 
Ω
1.1 Vrms
20.00 kHz
reference
40 kHz
60 kHz
80 kHz
100 kHz
> 70 dB
O
Sine wave
50 
Ω
1.1 Vrms
100.00 kHz
reference
200 kHz
300 kHz
400 kHz 500 kHz
> 60 dB
O
Sine wave
50 
Ω
1.1 Vrms
1.00 MHz
reference
2 MHz
3 MHz
4 MHz
5 MHz
> 45 dB
O
Sine wave
50 
Ω
1.1 Vrms
15.00 MHz
reference
30 MHz
45 MHz
60 MHz
75 MHz
> 35 dB
a.The E1441A has a fixed output impedance of 50 
Ω
 on the "Output" terminal.  Use the OUTPut:LOAD 50 | INFinity 
| MIN | MAX command to set the output termination. HIGH Z assumes no load on output. 50 
Ω
 assumes a 50 
Ω
 
±
 0.1 
Ω
 load on output.