VXi VX1410A User Manual

Page of 360
STATus Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–89
STATus:QUEStionable:TEMPerature:LIMit
STATus:QUEStionable:TEMPerature:LIMit?
Sets or queries the maximum allowable ambient and delta temperatures in
each slot. The factory default settings are 55 
°
C for ambient and 10 
°
C for each
delta temperature.
         
        
<Location>
Query response
{AMBient | DELTa0 . . . DELTa12}
AMBient - ambient input temperature
DELTa0 . . . DELTa12 – output temperature minus
input temperature of each slot
<NR1>
<Limit>
<NRf>
If Location is AMBient
 Limit 
 75
MIN
MAX
If Location is DELTan
 Limit 
 30
MIN
MAX
1
75
5
30
0
–224, “Illegal Parameter Value”
You attempted to set the location to an illegal value.
–222, “Data Out of Range”
You attempted to set the temperature limit to an illegal value.
None
Query:
      
Response:

STATus:QUEStionable:TEMPerature?
STATus:QUEStionable:TEMPerature:CONDition?
STATus:QUEStionable:TEMPerature:ENABle
STATus:QUEStionable:TEMPerature:LEVel?
Syntax
Reset Value
Errors and Events
Dependencies
Examples
Related Commands