Operating the Analyzer 
 
 
24 
UDA2182 Universal Dual Analyzer Product Manual 
January 2009 
 Online Functions 
Table 5-2 Display Details Functions 
Detail 
Function 
Process Variable 
Values 
When two input boards are installed, the default online screen displays both PVs and 
their units of measure, as determined by the input boards, the probe (if memory-
embedded) or any measurement configuration options that may be available. When 
only one input board is installed, the default online screen displays one PV and its units 
in a larger font size. 
The currently selected PV type determines the numerical format and the units of 
measure on the online PV display. Measured PV is generally displayed in the highest 
decimal precision possible with five digits and has a potentially displayable range of 
0.0000 to 99999. The exceptions are dissolved oxygen, pH, ORP and temperature, 
which are displayed with fixed decimal precision. 
PV Type determines specific ranges and in the case of Conductivity, cell constant 
determines available PV Types. Each PV measurement and display is updated every 
500ms maximum. Each temperature measurement and display is updated every 10 
seconds maximum. 
See the Specific Input configuration for available ranges. (Section 6.6)
 
Tag Name 
The real-time displays of process values show the instrument’s tag name (or other 
configurable fixed sixteen-character string) at the top of the screen. 
PV Temperature 
Each PV value is accompanied by a temperature value for all measurements except 
ORP, as ORP probes do not contain temperature sensors and no measurement 
compensation for temperature is required. Temperature values are displayed in units of 
degrees Fahrenheit or degrees Celsius as determined by configuration. 
Measured temperature is always expressed in fixed tenths decimal precision and has a 
displayed range according to input type: 
PH/ORP 
-10.0 to 110.0
°C or 14.0 to 230.0°F 
Conductivity 
0 to 140.0
°C or 32.0 to 284°F 
Dissolved Oxygen 
0 to 60.0
°C or 32 to 140°F 
 
Status Messages 
A text string appears on the bottom of all displays. Online displays provide messages 
relaying online diagnostics, alarms and other events. Offline screens display messages 
relevant to data entry and calibration. See Section 12. 
Bargraphs 
The Bargraphs will represent up to three output values. The corner indicators represent 
the physical state of the Relay Outputs [1, 2, 3, and 4].  
*Note that all values and indicators on the main (input) display screen are maintained in the input setup 
group.