Agilent Technologies 1660AS User Manual
NWIDth?
Query
:MEASure:[SOURce CHANnel<N>;]NWIDth?
The NWIDth query makes a negative width time measurement on the
selected channel. The measurement is made between the 50% points of the
first falling and the next rising edge displayed on screen.
selected channel. The measurement is made between the 50% points of the
first falling and the next rising edge displayed on screen.
Returned Format
[:MEASure:NWIDth] <value><NL>
<N>
An integer from 1 to 2
<value>
negative pulse width in seconds
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN2;NWID?"
OVERshoot?
Query
:MEASure:[SOURce CHANnel<N>;]OVERshoot?
The OVERshoot query makes an overshoot measurement on the selected
channel. The measurement is made by finding a distortion following the first
major transition. The result is the ratio of OVERshoot vs. VAMPlitude.
channel. The measurement is made by finding a distortion following the first
major transition. The result is the ratio of OVERshoot vs. VAMPlitude.
Returned Format
[:MEASure:OVERshoot]<value><NL>
<N>
An integer from 1 to 2
<value>
ratio of overshoot to Vamplitude
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN1;OVER?"
MEASure Subsystem
NWIDth?
32-7