Invacare MA95Z User Manual
SECTION 2—OPERATION
18
Mode Button (MA90Z)
1. Press the Mode button (
) to select the Max Inflate Mode or the Low Air Loss mode.
• Max Inflate Mode ‐ In this mode, the mattress inflates rapidly to maximum
firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three
minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the
Max Inflate mode deactivates and the control unit defaults to the previous setting.
minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the
Max Inflate mode deactivates and the control unit defaults to the previous setting.
NOTE: It takes 5‐10 minutes for the MA90Z mattress to inflate fully (inflation time depends
on size of mattress).
on size of mattress).
NOTE: Max Inflate mode can be manually disengaged by pressing the Max Inflate button.
This will deactivate the Max Inflate LED.
This will deactivate the Max Inflate LED.
NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress,
patient wound care, patient turning or patient cleaning.
patient wound care, patient turning or patient cleaning.
• Low Air Loss Mode ‐ In this mode, the mattress goes into on‐demand low air loss
relief mode.
Max Inflate Button (MA95Z)
1. Press the Max Inflate button (
) to select the max inflate mode and inflate the
mattress rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of
beeps sound every three minutes as a reminder that the Max Inflate mode is active.
After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the
previous setting.
beeps sound every three minutes as a reminder that the Max Inflate mode is active.
After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the
previous setting.
NOTE: It takes 25‐60 seconds for the MA95Z mattress to inflate fully (inflation time depends on
size of mattress).
size of mattress).
NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, patient
wound care, patient turning or patient cleaning.
wound care, patient turning or patient cleaning.
2. Perform one of the following:
• MA90Z ‐ Press the Max Inflate button again to manually disengage Max Inflate
mode.
• MA95Z ‐ Select a different setting to exit Max Inflate mode.
NOTE: This will deactivate the Max Inflate LED.