SECTION 3—MAINTENANCE AND TROUBLESHOOTING
25
Changing the Filter
NOTE: For this procedure, refer to FIGURE 3.1 on page 25.
Clean the filter every 5 months or whenever it is dirty. 
1. Remove the two screws securing the filter cover to the back of the control unit.
2. Remove the filter.
3. Wash the filter with water and a mild detergent.
4. Dry the filter completely.
5. Install the filter into the back of the control unit.
6. Secure the filter cover to the back of the control unit using the two screws.
FIGURE 3.1   Changing the Filter
Troubleshooting
PROBLEM
CAUSE
SOLUTION
Mattress not inflating
Not alternating properly
Mattress hose disconnected
Connect hose connectors and lock 
them in place
Air hose kinked or split
Unkink hose or replace split hose
Major leak in the air cushions or 
overlay pad
Replace leaking air cushions or 
overlay pad
Kinked or split manifold
Unkink manifold or replace split 
manifold
Has power and fuse is good, control 
unit does not come on
Send unit back for repair
Not alternating, solenoid or valve 
malfunction
Send control unit back to factory for 
repair
No air, pump or blower malfunction. Send unit for repair