Invacare MA55 User Manual

Page of 24
SECTION 1—GENERAL GUIDELINES
9
Fire Hazard
 DANGER
Smoking
DO NOT SMOKE while using this device. This system uses room air for circulation 
through the mattress. A cigarette can burn a hole in the bed surface and cause 
damage to the mattress. Also, patient clothing, bed sheets, etc. may be combustible 
and cause a fire. Failure to observe this warning can result in severe fire, property 
damage and cause physical injury or death.
Smoking by visitors in the room will contaminate the system. Therefore, visitor 
smoking is NOT permitted.
Anesthesia Equipment
There is an explosion risk if used with flammable anesthetics. 
Oxygen
There is a possible fire hazard when used with oxygen administering equipment 
other than nasal mask or half bed tent type. The oxygen tent should NOT extend 
below mattress support level.
Electrical
 DANGER
Electrical shock hazard. DO NOT remove cover. Refer to qualified service 
personnel. 
Before performing any maintenance to the power unit, disconnect the power cord from the 
wall outlet.
DO NOT insert items into any openings of the control unit. Doing so may cause fire or 
electric shock by shorting the internal components.
The control unit MUST be kept away from all heat sources and radiators during operation.
Connect the equipment to properly grounded three prong wall outlet using 10-14 ft hospital 
grade power cord provided with the product.
Grounding reliability depends upon a properly grounded receptacle (3-prong).