51, 52, 53, 54 Series II
Service Manual
8
Table 3. Performance Test Values (K-Type Thermocouple)
Input
Voltage
Equivalent
Display Reading limits
Step
(mV DC)
Temperature
°
C
°
F
1
0.0000
0.0 
°
C
32.0 
°
F
-0.3  to  0.3
31.5  to  32.5
2
0.9193
23.0 
°
C
73.4 
°
F
22.7  to  23.3
72.9 to 73.9
3
-5.7297
-190.0 
°
C
-310.0 
°
F
-189.3  to  -190.7
-308.9  to  -311.1
4
40.8853
990.0 
°
C
1814 
°
F
989.2 to  990.8
1813  to 1815
5
48.8382
1200 
°
C
2192 
°
F
1199  to 1201
2190 to 2194
Verification of IR Communications Port (Fluke 53 and 54)
This section explains how to verify the operation of the IR Communications Port (IR
port). Perform the following procedure using a PC and HyperTerminal. The IR port can
also be verified using Flukeview
 software.
To set up a HyperTerminal file, follow these steps:
1.
 
Connect the IR adapter cable to the DB9 Adapter connector. Connect the other end of
the DB9 Adapter to a COM port of the PC.
2.
 
Place the Fluke 53 or 54 in front of the IR adapter.
3.
 
Open up a HyperTerminal session. Select 
Start button
Programs
Accessories
,
HyperTerminal
.
4.
 
Open up a 
New Connection
 with a name like “Fluke 54 IR Test”.
5.
 
When the 
Connect To
 box appears, select the COM1 or COM2 (the port that the IR
adapter cable is plugged into).
6.
 
Select 
Port Settings
: 9600 baud rate, 8 data bits, no parity, 1 stop bit, and no flow
control.
7.
 
Under 
File
, select 
properties
. Verify that the correct port is selected. Click on
Configure
. Verify that port settings are correct (the baud rate may need to be
changed to 9600).
8.
 
Select the 
Settings
 tab then click on 
ASCII Setup
. Check the following boxes:
 
Echo typed characters locally
 
Append line feeds to incoming line ends
 
Wrap lines that exceed terminal width
To open a HyperTerminal file to test the IR port, use the following procedure:
1.
 
Under 
File
, select 
Save As
. Use a file name such as “Fluke 54 IR Test” and then
click on 
save
.
2.
 
Click on the 
phone disconnect
 icon.
3.
 
Click on 
open folder
 or under 
File
 select 
Open
. Double click on Fluke 54 IR Test.
4.
 
Type in "ID" followed by the 
Enter
 key (notice the "ID" is capitalized).
5.
 
The display of the thermometer now shows the thermometer name (Fluke 53-II or
Fluke 54-II), and the units operating software version (e.g. Fluke 54-II, V1.01).