Fluke 54 User Manual

Page of 28
Thermometer
Performance Test and Calibration Required Test Equipment
5
Performance Test and Calibration Required Test Equipment
Refer to Table 2 for a list of test equipment used in the performance test and calibration
procedures.
Table 2. Recommended Test Equipment
Equipment
Required Characteristics
Recommended Model
DC Voltage Calibrator
Output Voltage:  0-1V
Accuracy: 0.0125 %
Resolution: 0.1 
µ
V
Fluke 5700A or equivalent
Thermocouple connector
For K-type thermocouple
Fluke 80CK-M
Fluke P/N 779942
Thermocouple connector
Copper
Fluke P/N 601747
K-type Thermocouple wire
Standard
Omega
Dewar Flask/Cap
1- pint, or larger capacity
Thermos
Mercury Thermometer
(either 
°
C or 
°
F)
0.02 
°
C Resolution
0.05 
°
F Resolution
Princo Model ASTM56C
Princo Model ASTM56F
DB9 Adapter Connector
Fluke P/N 804549
IR Adapter
Fluke P/N 690518
PC with Microsoft Windows
®
installed
Performance Tests
Performance tests are recommended for incoming inspection, periodic maintenance, and
for checking the thermometer’s specifications. If the thermometer fails any part of the
performance tests, calibration and/or repair is indicated.
Before performing these tests:
1.
 
Allow the thermometer to stabilize to room temperature, +23 
o
±
 5 
o
C (+73 
o
±
 9 
o
F).
2.
 
Check the batteries and replace them if necessary.
Display Test
Turn the thermometer on while pressing the 
h
 button. All display segments are
displayed until the 
h
 button is released. Confirm that all the display segments are
showing as in Figure 1.