Leica CM3050 S User Manual

Page of 66
10
Instruction Manual V1.2 - 11/2006
Instruments with sectioning motor:
The sectioning motor is now blocked.
All instruments:
The message ‘LOCKED’ in the display of control
panel 1 indicates that the handwheel has been
locked:
Unlocking:
To unlock, push locking pin (2) to the left into
position (2a).
Display indication ‘LOCKED’ disappears.
Instruments with sectioning motor:
The sectioning motor can now be activated
again.
2.2.2 Centering the handwheel grip
During motorized sectioning, for
safety reasons always center the
handwheel grip!
To center grip (1), pull outwards and pivot
into center of handwheel.
When released, the grip locks into position.
2.2.1 Locking the handwheel
Always cover the cutting edge with
the knife guard and lock the hand-
wheel:
• Prior to doing any work on
knife and/or specimen.
• Prior to exchanging specimens.
• During work breaks.
The handwheel can be locked in 2 positions:
with the grip in the uppermost position (left),
with the grip in the lowest position (right).
Locking:
Rotate handwheel, until grip (1) is in upper or
lower position.
To lock, press pin (2) to the right into position
(2b).
--> The upper locking position for pin (2) is
marked by a black dot (4).
2.
Safety
2
2
C T - 3 0 ° C
O T - 3 5 ° C
L O C K E D