Maintenance instructions
General Notes
 
Ϙ
Protect the microscope from moisture, 
fumes and acids and from alkaline, caustic 
and corrosive materials. Do not store 
chemicals in the vicinity.
 
Ϙ
Protect the microscope from oil and grease. 
Never grease or oil mechanical parts or 
sliding surfaces.
 
Ϙ
Follow the instructions of the disinfectant 
manufacturer.
 
Ϙ
We recommend concluding a service 
contract with Leica Service.
Cleaning coated Parts and Plastic Parts
 
Ϙ
Dust and dirt particles should be removed 
with a soft brush or lint-free cotton cloth.
 
Ϙ
Remove coarse debris with a moistened 
disposable cloth.
 
Ϙ
Acetone, xylene or nitro-containing thin-
ners must NOT be used.
 
Ϙ
Never use chemicals to clean colored 
surfaces or accessories with rubberized 
parts. This could damage the surfaces, and 
specimens could be contaminated by abra-
ded particles.