Tektronix 070-9774-02 User Manual

Page of 82
Basic Installation
10
TLA 700 Series Installation Manual
Power On the Logic Analyzer
Follow these steps to power on the logic analyzer for the first time.
1. Check that the line fuse is correct for your application. See Table 3 and
Figure 6.
Table 3: Line fuses 
Line voltage
Rating
Tektronix part number
Portable mainframe
90 V to 132 V operation
8 A, fast blow, 250 V
159-0046-00
207 V to 250 V operation
6.3 A, fast blow, 250 V
159-0381-00
Benchtop mainframe
90 V to 132 V operation
20 A, slow blow, 250 V
159-0379-00
103 V to 250 V operation
15 A, fast blow, 250 V
159-0256-00
207 V to 250 V operation
6.3 A, fast blow, 250 V
159-0381-00
AC power and fuse
Figure 6: Line fuse and power cord connector locations
CAUTION. Connect the keyboard, mouse, and other accessories before applying
power to the mainframe. Connecting the accessories after powering on the
mainframe can damage the accessories. Refer to Figure 4 on page 7 for
connection information.