Fluke 51 & 52 Series II User Manual

Page of 20
51 & 52 Series II
Users Manual
14
General
Weight
280 g (10 oz)
Dimensions
(without
holster)
2.8 cm 
×
 7.8 cm 
×
 16.2 cm
(1.1 in 
×
 3 in 
×
 6.4 in)
Battery
3 AA batteries
Certification
P
)
 (TÜV pending)
Safety
CSA C22.2 No. 1010.1 1992
EN 61010 Amendments 1, 2
CAT I
OVERVOLTAGE (Installation)
CATEGORY I, Pollution Degree 2
per IEC1010-1*
Refers to the level of Impulse Withstand Voltage protection
provided. Equipment of OVERVOLTAGE CATEGORY I is
equipment for connection to circuits in which measures are
taken to limit the transient over voltages to an appropriate low
level. Example include protect electronic circuits.
80 PK-1 Thermocouple (supplied with
thermometer)
Type
Type K, Chromel Alumel, bead style
Temperature
Range
40 oC to 
+
260 oC
(
40 oF to 
+
500 oF)
Accuracy
±
 1.1 oC (
±
 2.0 oF)
Electrical
Measurement
Range
J-type: 
210 oC to 
+
1200 oC
   (
346 oF to 
+
 2192 oF)
K-type: 
200 oC to 
+
1372 oC
   (
328 oF to 
+
2501 oF)
T-type: 
250  oC to 
+
400 oC
   (
418 oF to 
+
752 oF)
E-type: 
150  oC to 
+
1000 oC
   (
238 oF to 
+
1832 oF)
Display
Resolution
0.1  oC / oF / K 
<
 1000o
1.0 oC / oF / K 
 1000o