DC Voltage
±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range
1281
HP3458/HFL
8508A
120.00000m V
-
4 + 3
-
199.999999m V
6 + 0.5
-
2.7 + 0.5
1.20000000V
-
3 + 0.3
-
1.99999999V
3 + 0.2
-
2.7 + 0.2
12.000000V
-
3 + 0.05
-
19.9999999V
3 + 0.2
-
2.7 + 0.2
120.00000V
-
5 + 0.3
-
199.999999V
6 + 0.1
-
4.0 + 0.2
1050.0000V
6 + 0.1
5 + 0.1
4.0 + 0.5