15. Click the Upload button to start the firmware upgrade.
16. When upload is finished, select Diagnostics in the Tools menu.
17. Click Device Errors and check for “Checksum”.
18. If it is on the list, choose the Factory Settings option from the Tools 
menu.
Checksum