d. Gas 
Connections 
Sample In ...................................... 1/4" O.D. tube fitting, stainless steel 
Regulated Air In ...............................1/4" O.D. tube fitting, brass 
Burner Air In .....................................1/4" O.D. tube fitting, brass 
Fuel In...............................................1/4" O.D. tube fitting, stainless steel 
Purge Air In ......................................3/8" O.D. tube fitting, brass 
Purge Air Out....................................3/8" O.D. tube fitting, brass 
Bypass Out.......................................1/4" O.D. tube fitting, stainless steel 
Burner Exhaust Out .........................3/8" O.D. tube connection, stainless steel (must slope downward 
6
o
 min. from horizontal) 
 
Note 
Burner Exhaust, Bypass Out and Purge Air Out to be vented to atmospheric pressure and to non-
classified location in accordance with ANSI/NFPA-496 guideline 
 
Pressure Relief Valve .................... See Caution below 
 
CAUTION 
PRESSURE RELIEF VALVE 
No connection shall be made to this fitting.  If this caution is ignored, damage to the case seals 
could occur, and the instrument will not operate properly 
 
WARNING 
HIGH TEMPERATURE 
The Sample In, Bypass Out, and Burner Exhaust Out connections can reach temperatures of up to 
250
°C (480°F).  Severe burns could result from touching these connections. 
 
 
 
 
 
 
See the Preface section of the Platform Components manual for specifications regarding Platform-