Measuring and Displaying Parameters
Specifications
12-7
NBP (Non-
Invasive
Blood
Pressure)
Temperature
Parameter
Specification
Standards
Meets performance standards of ANSI/
AAMI SP10-1992
AAMI SP10-1992
Technique
Oscillometric
Cuff Pressure Display
10 to 300 mmHg
Pulse Rate Range
40 to 200 b/min
Blood Pressure Accuracy
Mean error and standard deviation per
ANSI/AAMI SP10-1992
ANSI/AAMI SP10-1992
a
a. Systolic and Diastolic blood pressure measurements determined with this device are
equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation
method, and mean pressure measurements determined with this device are equivalent to
those obtained by an intra-arterial blood pressure measurement device, within the limits
prescribed by the American National Standard, Electronic or automated
sphygmomanometers.
method, and mean pressure measurements determined with this device are equivalent to
those obtained by an intra-arterial blood pressure measurement device, within the limits
prescribed by the American National Standard, Electronic or automated
sphygmomanometers.
Pulse Rate Accuracy
Greater of ± 2 BMP or ± 2%
Initial Cuff Inflation
180 mmHg
Subsequence Cuff Inflation
Prev SYS ±30 mmHg
Blood Pressure Measurement Range
Systolic
60 to 250 mmHg
Diastolic
40 to 220 mmHg
Mean Arterial Pressure
45 to 235 mmHg
Measurement Modes
Auto
Automatic measurements at intervals of
1, 3, 5, 10, 30, 60 and 90 minutes
1, 3, 5, 10, 30, 60 and 90 minutes
Manual
Single measurement initiated by Start/
Stop button
Stop button
STAT
Series of consecutive measurements for
5 minutes
5 minutes
Parameter
Specification
Technique
Probe
Range
15 to 45
o
C (59 - 113
o
F)
Accuracy
± 0.1
o
C