Setting up Spirometry
15 Monitoring Airway Flow, Volume and Pressure
193
WARNING
Safety and effectiveness of the respiration measurement method in the detection of apnea, particularly
the apnea of prematurity and apnea of infancy, has not been established.
the apnea of prematurity and apnea of infancy, has not been established.
WARNING
If an apnea occurs during a zero calibration, the time delay between the start of apnea and the
activation of the apnea alarm could be up to 10 seconds plus the configured apnea delay time.
activation of the apnea alarm could be up to 10 seconds plus the configured apnea delay time.
Automatic Alarm Suppression
If the automatic alarm suppression is switched on, all spirometry alarms are suppressed until breathing
activity has been detected. This setting must be made in configuration mode.
activity has been detected. This setting must be made in configuration mode.
Choosing Measured AWV Components
1
In the
Setup Spirometry
menu, select
AWV
.
2
Select
MV
to choose the measured components for Minute Volume
(inspiratory and/or expiratory or off ).
Select
Select
TV
to choose the measured components for Tidal Volume
(inspiratory and/or expiratory or off ).
N O T E
If MV and TV are set to off, there will be no alarming for these parameters.
Setting the Gas Compensation Mode
1
In the
Setup Spirometry
menu, select
Gas Compensation
.
2
Select
Mode
to choose the Gas Compensation mode:
– select
Manual
for manually entered gas concentrations or
– select
Gas Analyzer
for gas concentrations from the gas analyzer.
N O T E
• Gas concentrations from the gas analyzer are only available for Philips gas analyzers, not for devices
connected via the Vuelink module.
• If gas concentrations from the gas analyzer are selected but not all data is available, the missing data
is taken from manually entered values. In case of invalid data or no data at all, the INOP message
SPIRO GAS COMPENS? is displayed.
SPIRO GAS COMPENS? is displayed.
Changing the Type of Balance Gas
1
In the
Setup Spirometry
menu, select
Gas Compensation
.
2
Select
Balance Gas
to change the type of balance gas.
3
Select the appropriate setting (N
2
or N
2
O).
Changing the Concentration of Inspired O
2
and the Inspired Agents
1
In the
Setup Spirometry
menu, select
Gas Compensation
.
2
Select
Inspired O2
to change the concentration of inspired O
2
.
Select
Inspired Agent
to change the concentration of the inspired agent(s).