Setting up Spirometry
15 Monitoring Airway Flow, Volume and Pressure
193
WARNING
Safety and effectiveness of the respiration measurement method in the detection of apnea, particularly 
the apnea of prematurity and apnea of infancy, has not been established.
WARNING
If an apnea occurs during a zero calibration, the time delay between the start of apnea and the 
activation of the apnea alarm could be up to 10 seconds plus the configured apnea delay time. 
Automatic Alarm Suppression
If the automatic alarm suppression is switched on, all spirometry alarms are suppressed until breathing 
activity has been detected. This setting must be made in configuration mode.
Choosing Measured AWV Components
1
In the 
Setup Spirometry 
menu, select 
AWV
.
2
Select 
MV
 to choose the measured components for Minute Volume
(inspiratory and/or expiratory or off ).
Select 
TV
 to choose the measured components for Tidal Volume
(inspiratory and/or expiratory or off ).
N O T E
If MV and TV are set to off, there will be no alarming for these parameters.
Setting the Gas Compensation Mode
1
In the 
Setup Spirometry 
menu, select 
Gas Compensation
.
2
Select 
Mode 
to choose the Gas Compensation mode:
– select 
Manual
 for manually entered gas concentrations or
– select 
Gas Analyzer
 for gas concentrations from the gas analyzer.
N O T E
• Gas concentrations from the gas analyzer are only available for Philips gas analyzers, not for devices 
connected via the Vuelink module.
• If gas concentrations from the gas analyzer are selected but not all data is available, the missing data 
is taken from manually entered values. In case of invalid data or no data at all, the INOP message 
SPIRO GAS COMPENS? is displayed.
Changing the Type of Balance Gas
1
In the 
Setup Spirometry 
menu, select 
Gas Compensation
.
2
Select 
Balance Gas 
to change the type of balance gas.
3
Select the appropriate setting (N
2
 or N
2
O).
Changing the Concentration of Inspired O
2
 and the Inspired Agents
1
In the 
Setup Spirometry 
menu, select 
Gas Compensation
.
2
Select 
Inspired O2 
to change the concentration of inspired O
2
.
Select 
Inspired Agent 
to change the concentration of the inspired agent(s).