Battery Power Indicators
31 Using the Batteries
299
Battery status symbols:  These symbols tell you the status of the batteries (see table below) and which 
battery compartment they are in, either 1 or 2. 
Battery power gauge:  This shows the remaining battery power in the combined batteries. It is divided 
into sections, each representing 20% of the total power. If three and a half sections are shaded, as in 
this example, this indicates that 70% battery power remains. If no batteries are detected, the battery 
gauge is grayed-out. If no data is available from the batteries, questions marks are shown in the gauge.
Monitoring Time Available: Below the battery power gauge a time is displayed. This is the estimated 
monitoring time available with the current battery power. Note that this time fluctuates depending on 
the system load (the display brightness and how many measurements and recordings you carry out), 
the age of the battery, and the remaining capacity of the battery. 
Battery malfunction symbols:  If a problem is detected with a battery, these symbols alternate with 
the symbol indicating the position of the battery affected. They may be accompanied by an INOP 
message or by a battery status message in the monitor information line providing more details. Symbols 
indicating critical situations are colored red.
If both batteries are malfunctioning or incompatible, the monitor will switch off automatically for 
safety reasons, if it is not connected to AC power.
Battery status symbols
Battery malfunction symbols
Battery 1 is 
present
Battery 
compartments 
are empty
Battery requires 
maintenance
Incompatible 
battery 
(Battery 1)
Battery 
malfunction 
(Battery 1)
Battery 2 is 
missing, insert 
battery (MP40/
50 only)
battery is 
empty
battery not 
charging as the 
temperature is 
above or below 
the specified 
range
(red) battery 
temperature 
too high
Battery 2 has no 
power left 
1
2
2
1
1