4 Patient Alarms and INOPs
Patient Alarm Messages
54
***BRADY (Pulse) 
or 
***BRADY xxx<yyy
Press, SpO
2
The heart rate from the Pulse signal has 
fallen below the bradycardia limit. xxx 
denotes the lowest measured value; yyy 
is the bradycardia limit.
numeric flashes and alarm 
limit is highlighted, red alarm 
lamp, alarm tone.
**CCO/CCI HIGH
CCO
Continuous Cardiac Output or CC 
Index is above the high alarm limit.
numeric flashes and high alarm 
limit is highlighted, yellow 
alarm lamp, alarm tone.
**CCO/CCI LOW
CCO
Continuous Cardiac Output or CC 
Index is below the low alarm limit.
numeric flashes and low alarm 
limit is highlighted, yellow 
alarm lamp, alarm tone.
**CPP HIGH
CPP
The CPP value has exceeded the high 
alarm limit.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone
**CPP LOW
CPP
The CPP value has fallen below the low 
alarm limit.
numeric flashes and low limit 
is highlighted, yellow alarm 
lamp, alarm tone.
***DESAT 
or 
***DESAT xxx<yyy
SpO
2
The SpO
2
 value has fallen below the 
desaturation alarm limit. xxx denotes 
the lowest measured value, and yyy is 
the desaturation limit. 
numeric flashes, red alarm 
lamp, alarm tone.
**et <Agent label> 
HIGH
AGM 
The end tidal agent high alarm limit has 
been exceeded.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**et <Agent label> 
LOW
AGM 
The end tidal agent value has fallen 
below the low alarm limit.
numeric flashes and low alarm 
limit is highlighted, yellow 
alarm lamp, alarm tone.
**etCO2 HIGH
CO
2
, Resp, 
AGM
The end tidal CO
2
 high alarm limit has 
been exceeded. 
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**etCO2 LOW
CO
2
, Resp, 
AGM
The end tidal CO
2
 value has fallen 
below the low alarm limit. 
numeric flashes and low limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**etO2 HIGH
O
2
, AGM
The end tidal O
2
 high alarm limit has 
been exceeded.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**etO2 LOW
O
2
, AGM
The end tidal O
2
 value has fallen below 
the low alarm limit. 
numeric flashes, and low limit 
is highlighted, yellow alarm 
lamp, alarm tone.
*/**/*** 
EVENT:<GRP> 
Event 
surveillance
An event has occurred and the event 
notification is configured to alarm. 
<GRP> is the event group
event group name flashes, 
yellow or red alarm lamp and 
alarm tone
*/**/*** EVENT
at Information center
Event 
surveillance
An event has occurred and the event 
notification is configured to alarm. 
Check on the monitor for more details 
on event group.
(on monitor) event group 
name flashes, yellow or red 
alarm lamp and alarm tone
***EXTREME BRADY
ECG
The bradycardia limit has been 
exceeded.
numeric flashes and alarm 
limit is highlighted, red alarm 
lamp, alarm tone.
***EXTREME TACHY
ECG
The tachycardia limit has been 
exceeded.
numeric flashes and alarm 
limit is highlighted, red alarm 
lamp, alarm tone.
Alarm Message
From
Condition
Indication