Philips MP40/50 Manual De Usuario

Descargar
Página de 414
25 Printing Patient Reports
Sample Report Printouts
276
ECG Reports
Below the header on ECG Reports, the numeric block shows the current HR, PVC, and ST values. 
The wave area shows the printed waves for all available ECG leads. A 1 mV calibration bar is printed at 
the beginning of each wave. With the 3X4, 6X2, and 2X4 formats, a rhythm stripe prints a longer 
section of the ECG wave from the primary ECG lead for ECG rhythm evaluation. The ECG signal 
bandwidth, the patient’s paced status, the ECG gain, and the print speed are printed at the bottom of 
the wave area. Pace pulse marks are automatically printed beside the wave for paced patients. Beat 
labels can be set to print on the rhythm stripe. The 12X1 (2P) format prints the report over two pages.
Other Reports
See the sections on Trends and Calculations and the chapter on Event Surveillance for other example 
reports.
ECG Report type
Available Formats
Available Paper Sizes
12-Lead ECG
3X4 landscape
A4, letter, A3, ledger
6X2 landscape
A4, letter, A3, ledger
12X1 portrait
A4 and letter only
12X1 landscape
A4, letter, A3, ledger
12X1 (2P) landscape
A4, letter (2 pages)
Multi-lead ECG
2X4 landscape
A4, letter, A3, ledger
8X1 portrait
A4 and letter only
8X1 landscape
A4, letter, A3, ledger
Patient 
information
Numeric 
block
Wave area