Philips MP20/30 Manual De Usuario

Descargar
Página de 414
203
17
17
Monitoring SvO
2
MP60/70/
80/90
monitors
only
The SvO
2
 module measures the percentage of mixed venous oxygen saturation continuously and 
invasively using the Hospira, Inc. (formerly Abbott Laboratories) OptiCath family of catheters routed 
via the right side of the heart into the pulmonary artery. Can be used only with the MP60/MP70/
MP80/MP90 monitors. 
WARNING
Injected dyes, such as methylene blue, or intravascular dyshemoglobin may lead to inaccurate 
measurements.
Do not monitor oxygen saturation during infusion of I.V. fat emulsion or other turbid substances 
through the distal lumen of the OptiCath catheter. These liquids might temporarily modify the blood 
scattering and absorption characteristics at the catheter tip. This interferes with the optical 
measurement of oxygen saturation. After infusion is complete, you can again monitor oxygen 
saturation accurately.
During injection of the bolus for thermodilution cardiac output measurements, the SvO
measurement might be disturbed.