19 Monitoring BIS
BIS Continuous Impedance Check
218
Manufacturer’s Information
BISx, the BIS Engine, the DSC, the Patient Interface Cable and the BIS Sensors are manufactured by 
Aspect Medical Systems.
BIS Continuous Impedance Check
This checks: 
• the combined impedance of the signal electrodes plus the reference electrode. 
This is done continuously and does not affect the EEG wave. As long as the impedances are within 
the valid range, there is no notification of this check or its results. 
• the impedance of the ground electrode. 
This is done every ten minutes and takes approximately 4 seconds. It causes an artifact in the EEG 
wave, and the message 
Ground Check
 is shown on the monitor screen during the check. If the 
ground electrode does not pass this check, another check is started. This continues until the ground 
electrode passes the check.
If the Continuous Impedance Check interferes with other measurements, it can be switched off. To do 
this:
1
In the 
Setup BIS
 menu, select
 Cont.Imp.Check.
2
Select 
On
/
Off
 to toggle between the settings.
CAUTION
Switching the continuous impedance check off will disable automatic notification to the user of 
impedance value changes, which may lead to incorrect BIS values. Therefore, this should only be done 
if the check interferes with or disturbs other measurements.
When a Semi-reusable Sensor is in use the continuous impedance check cannot be switched off. If it is 
switched off when the sensor is connected, it will automatically be switched on again.
BIS Cyclic Impedance Check 
This measures the exact impedance of each individual electrode. It causes a disturbed EEG wave, and 
the INOP 
BIS IMPEDANCE CHCK
 is shown on the monitor screen during the check. 
Starting a Cyclic Impedance Check
The Cyclic Impedance Check is automatically started when a sensor is connected. To manually start a 
Cyclic Impedance Check: 
select 
Cyclic Check
 in the 
BIS Setup 
menu to toggle between On and Off, or 
select 
Start Cyclic Check
 in the BIS window.
Stopping a Cyclic Impedance Check
The Cyclic Impedance Check stops automatically if the impedances of all electrodes are within the 
valid range. To manually stop a Cyclic Impedance Check:
select 
Cyclic Check
 in the 
BIS Setup menu 
to toggle between On and Off, or