Emerson 2200S Manual De Usuario

Descargar
Página de 158
74
Micro Motion
®
 Model 2200S Transmitters
Testing and Moving to Production
Figure 7-2
Communicator – Flowmeter zero procedure
Figure 7-3
ProLink II – Flowmeter zero procedure
Modify zero time if 
required
Perform auto zero
Auto Zero Failed
Auto Zero Complete
OK
1 Auto zero
Calibration in Progress
message
OK
On-Line Menu >
     3 Diag/Service >
          3 Calibration
Troubleshoot
Modify zero time
if required
Calibration
Failure LED
Calibration in Progress
LED turns red
Green
Troubleshoot
Red
Perform Auto Zero
Done
ProLink >
     Calibration >
          Zero Calibration
Wait until Calibration in 
Progress LED turns green