Emerson DO-03 Manual De Usuario
MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
ITEM
CHOICES
FACTORY SETTINGS
CONFIGURE (continued)
D.Temperature compensation (Section 5.7)
1. Temperature compensation
1. Temperature compensation
Auto or manual
Auto
2. Manual temperature
-15 to 130°C
25°C
E. Noise Reduction (section 5.8)
Noise rejection
Noise rejection
50 or 60 Hz
60 Hz
F. Barometric Pressure (Section 5.9)
1. Barometric pressure
1. Barometric pressure
Auto or manual
Auto
2. Barometric pressure manual
0.4 - 1.2 atm (or equivalent units)
1.00 atm
3. Barometric pressure units
mm Hg, in Hg, bar, kPa
mm Hg
G. Main sensor calibration (Section 5.10)
1. Stabilize reading
1. Stabilize reading
a. oxygen (ppm)
0 to 20 ppm
0.05 ppm
b. oxygen (% saturation)
0 to 200%
1%
2. Stabilize time
0 - 30 sec
10 sec
3. Sensor zero stabilization value
4. Salinity (parts per thousand)
4. Salinity (parts per thousand)
0.0 - 99.9
o
/
oo
0.0
o
/
oo
H. Security (Section 5.11)
1. Lock all
1. Lock all
000-999 (000 disables)
000
2. Lock program
000-999 (000 disables)
000
3. Lock configuration
000-999 (000 disables)
000
TABLE 5-1. Program Settings List (continued)
23