SECTION 7—TROUBLESHOOTING GUIDE
26
SECTION 7—TROUBLESHOOTING 
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
SHORT BEEPS, 
LONG PAUSE
Concentrator 
NOT operating,
power switch ON.
Beep......Beep......
Main Power Loss:
1. Power cord NOT 
plugged in.
1. Insert plug into outlet.
2. No power at outlet.
2. Inspect house circuit break-
ers or fuses. If problem recurs, 
use a different outlet.
3. Tripped circuit breaker. 3. Call Service Provider.
Alarm:
CONTINUOUS
Concentrator 
NOT operating, 
Power Switch ON.
Beeeeeeeep....
System Failure:
1. Unit overheating due to 
blocked air intake.
1a. Remove and clean cabinet 
filters.
1b. Move oxygen concentrator 
at least three inches away from 
walls, draperies or furniture.
2. Insufficient power at 
outlet.
2. DO NOT use extension 
cords. Move to another elec-
trical outlet or circuit.
3. Internal repairs 
required
3. Call Service Provider.
Alarm:
RAPID
Beep....Beep...
Beep...Beep
Low Flow Alarm:
1. Kinked or blocked tub-
ing, cannula or humidifier.
1. Inspect for kinks or 
blockages. Correct, clean or 
replace item. Once 
corrected, turn power OFF 
for 60 seconds and then turn 
power back ON.
2. Flowmeter set too low: 
0.5 L/min (or less) for 
Platinum 5; 1.0 L/min (or 
less) for Platinum 10.
2. Reset flowmeter to pre-
scribed flowrate. If prescribed 
flowrate is below 0.5 L/min, 
use the Pediatric Flowmeter 
IRCPF16 (Only with Platinum 
5 and XL).