Siemens LR 460 Manual De Usuario

Descargar
Página de 123
7ML19985JM01
SITRANS LR 460 – INSTRUCTION MANUAL 
Page A-71
mmmmm
A
: T
e
c
hni
cal
 Ref
e
re
nce
Appendix A: Technical Reference
Principles of Operation
 
SITRANS LR 460 is a long range FMCW (Frequency Modulated Continuous Wave) radar 
transmitter. Radar level measurement uses the time of flight principle to determine 
distance to a material surface.
FMCW radar transmits a continuous wave. The frequency of the wave is constantly 
increasing: this is known as the sweep. By the time the first part of the wave has been 
reflected off the target and returned to the device, the part of the wave that is just being 
emitted is at a higher frequency. The difference in frequency between the transmitted and 
received signals is proportional to time of flight. 
Electromagnetic wave propagation is virtually unaffected by temperature or pressure 
changes, or by changes in the vapor or dust levels inside a vessel.
SITRANS LR 460 consists of an enclosed electronic component coupled to an antenna 
and process connection. The electronic component generates an electromagnetic signal 
from 24.2 GHz to 25.2 GHz that is directed to the antenna.
The signal is emitted from the antenna, and the reflected echoes are digitally converted 
to an echo profile. The profile is analyzed to determine the distance from the material 
surface to the reference point on the instrument. This distance is used as a basis for the 
display of material level and output.
Measurement Response
The measurement response (response rate) limits the maximum rate at which the 
display and output respond to changes in the measurement. Once the real process fill/
empty rate (m/s) is established, a response rate can be selected that is slightly higher 
than the application rate. The response rate automatically adjusts the filters that affect 
the output response rate. 
There are three preset options: slow, medium, and fast.
If none of the preset options is satisfactory, the filters can be adjusted individually.
Related 
Parameters
Response
Rate
LOE Timer 
(minutes)
Fill Rate
Empty Rate
Values
1 *
slow
100
 0.1 m/minute
 0.1 m/minute
2
medium
10
1 m/minute
1 m/minute
3
fast
1
10 m/minute
10 m/minute