MODEL 309 
FOUR CHANNEL DATALOGGING THERMOMETER 
 
 
4.0 Operating 
Instructions:
 
4.1 Preparation 
for 
Measurement 
Plug as many thermocouples as are to be monitored into the sockets marked 
T1 through 
T4, taking care to observe the polarity of the thermocouple pins. If a differential 
measurement is required, this is available only between 
T1 and T2, which determines 
which thermocouples should be connected to these locations. 
4.2  Power and Backlight 
The      button turns both the thermometer and the display backlight 
ON or OFF. Press it once to turn the thermometer ON. Every 
subsequent momentary press toggles the backlight ON or OFF. 
Press and hold this button for 3 seconds to turn the power OFF. 
When first powered on, the display will show how much memory 
space is available for use, as shown to the right. 
4.3  Selecting the Temperature Scale 
When first turned on, the instrument defaults to reading in Celsius (°C). The instrument 
toggles between Celsius (°C) and Fahrenheit (°F) each time the 
°C/°F button is pressed. 
The instrument remembers the scale setting when last turned off and powers on in that 
setting the next time. 
4.4 Data-Hold 
Operation 
The present reading is held on the display by pressing the 
HOLD button. Pressing the 
HOLD button again releases the hold function and returns the instrument to 
 
 
 
 
 
 
continuous reading.
 
  
In the 
HOLD mode, the T1-T2MAX MIN and °C/°F buttons are disabled, as indicated by 
two continuous beeps when pressed. 
4.5 T1-T2 
Operation 
When the 
T1-T2 button is pressed, the display will indicate 
1
 
 
indicating the instrument is in the differential mode of 
T1 minus T2
The display then indicates 
T1 value on the top left, T2 value on the 
bottom left, and the difference 
T1-T2 on the bottom right. 
4.6 Clock 
Setup 
To set the real time clock: 
 
1.  Press and hold the 
T1-T2 button and then power on the meter. 
 
2. Press 
HOLD (clock). 
 
3. Press 
REC  (
n) or °C/°F  (p) to increase or decrease the 
number; press 
HOLD to advance to the next item. The 
adjusting order is: year
; month; day; hour; minute. Press HOLD 
again after the last setting to complete the process.  
To abort during a setup process, press the       button. 
 
 
4.7  Recording Interval Setup 
To set the recording interval: 
 
1.  Press and hold the 
T1-T2 button and then power on the meter. 
 
2. Press 
MAX MIN (interval). 
 
3. Press 
REC  (
n) or °C/°F  (p) to increase or decrease the 
number; press 
MAX MIN to advance to the next item. Press 
MAX MIN again after the last setting to complete the process. 
To abort during a setup process, press the       button. 
 
MODEL 309 
FOUR CHANNEL DATALOGGING THERMOMETER 
 
 
TX
R X
G N D
4.8 Recording 
Data 
Each momentary press of the 
REC button will alternately start and 
stop recording. To clear the memory, power off the meter, press 
and hold the 
REC button followed by the       button, holding both 
on simultaneously for at least 5 seconds. The display will show 
‘CLR’ ‘SURE 5’, and releasing all buttons then clears the memory. 
4.9 MAX/MIN 
Operation 
Press the 
MAX MIN button to enter the MAX/MIN mode. In this mode, both the maximum 
and minimum values are simultaneously retained in memory and updated with every new 
data sample. The instrument first enters the MAX mode, and the MAX symbol and value 
are both displayed. Pressing 
MAX MIN again advances the display to the MIN symbol 
and value. The next press of the 
MAX MIN button will cause both the MAX and MIN 
symbols to flash. This indicates that the maximum and minimum values have been 
updated in memory and the displayed reading is the present temperature. Each 
successive press of the 
MAX MIN button circulates the display mode among these 
options. To exit the MAX/MIN mode, press and hold the 
MAX MIN  button  for     
two seconds. 
In the MAX/MIN mode, the 
°C/°F  button is disabled, as indicated by two consecutive 
beeps if pressed. 
4.10 Auto Power Off 
By default, the instrument powers on in the ‘Auto Power Off’ mode and will automatically 
shut off 30 minutes after the last key operation or RS232 communication.    
To disable this feature, press and hold the 
HOLD button and then power on the meter.  
Two successive beeps will indicate that ‘Auto Power Off’ is disabled, and the      will not 
be displayed.
 
4.11 Low Battery Condition  
When the battery voltage is at or below the minimum for proper operation, the    
 
symbol will show on the display indicating that the battery must be replaced. 
4.12 Calibration 
Calibration is conducted by adjusting the following potentiometers to within the 
tolerances shown. 
(Ambient Temperature: 23 ± 3°C)
 
Input Adjust 
VR 
Tolerance 
 
0 °C 
VR1 
 
± 0.1 °C 
 
190 °C 
VR2 
 
± 0.1 °C 
  1000 °C 
VR3 
 
± 1 °C 
  1900 °F 
VR4 
 
± 1 °F 
4.13 Digital Output 
The Digital Output is a 9600 bps N 81 serial interface. 
RX is a 5V normally high input port. 
TX is a 5V normally high output port. 
 
Appendix: Thermocouple Probe Specifications  
Model Range  Tolerances 
Description 
TP-K01 
Wire probe 
-50°C to 200°C 
-58°F to 392°F 
±2.2°C or ±0.75% 
±3.6°F or ±0.75% 
Teflon insulation. Maximum insulation 
temperature: 260°C (500°F) 
 
TP-K01: 
Probe for general condition measurements, especially 
for complex and hard to reach places.
 
 
A simple single point calibration may be 
performed by adjusting VR1 to 0 °C with 
thermally stabilized ice water.