Managing Climate Control
Designed to be ideal for most patients, Climate Control Auto enables the automatic delivery of a constant 
value of absolute humidity while protecting against rainout.
To allow for increased flexibility, Climate Control can be turned to Manual in the Setup menu. Setting 
Climate Control to Manual disables the automatic control of humidity and allows humidity and 
temperature levels to be set independently. However, rainout protection is not provided when Climate 
Control is set to Manual.
Mode
Humidity
Temperature
Setting range
Default settings
Setting range
Default settings
Climate Control – Auto
Climate Control
Constant 
absolute humidity 
(depending on 
temperature 
setting)
Off*, 60ºF–86ºF 
(16ºC–30ºC)
80ºF (27ºC)
Climate Control – 
Manual
Humidity level
Heated tube
Off–6.0 
(0.5 increments)
3
Off, 60ºF–86ºF 
(16ºC–30ºC)
80ºF (27ºC)
When the temperature setting is set to Off the tube will not heat the air, nor will the humidifier heat the water to 
add humidity to the air.