28
Sound:  
DECLARED 
DUAL-NUMBER 
NOISE 
EMISSION 
VALUES in 
accordance with 
ISO 4871:1996
Sound pressure level (CPAP mode)
with SlimLine air tubing: 26 dBA with uncertainty of 2 dBA as measured 
according to ISO 17510-1:2007
with Standard air tubing: 27 dBA with uncertainty of 2 dBA as measured 
according to ISO 17510-1:2007
with either SlimLine or 
Standard air tubing and H5i:
28 dBA with uncertainty of 2 dBA as measured 
according to ISO 17510-1:2007
Sound power level (CPAP mode)
with SlimLine air tubing: 34 dBA with uncertainty of 2 dBA as measured 
according to ISO 17510-1:2007
with Standard air tubing: 35 dBA with uncertainty of 2 dBA as measured 
according to ISO 17510-1:2007
with either SlimLine or 
Standard air tubing and H5i:
36 dBA with uncertainty of 2 dBA as measured 
according to ISO 17510-1:2007
Alarm volume 
settings
Low (nominal 56 dBA), Medium (nominal 68 dBA), High (nominal 80 dBA)
Physical
Dimensions (L x W x H): 6.0” x 6.8” x 3.4” (153 mm x 172 mm x 86 mm)
Weight: 2.30 lbs (1.045 kg)
Housing construction: Flame retardant engineering thermoplastic
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Air filter
Standard: Polyester non-woven fiber
Hypoallergenic: Acrylic and polypropylene fibers in a polypropylene carrier
Supplemental 
oxygen
Recommended maximum supplemental oxygen: 15 L/min (CPAP, S, T, ST, ASV and 
PAC modes), 4 L/min (AutoSet, VAuto, iVAPS and ASVAuto modes)
Pneumatic flow path
9
2
5
4
6
7
3
1
8
3
1.  Flow sensor
2.  Blower
3.  Pressure sensor
4.  Mask
5.  Air tubing
6.  H5i
7.  Alarm module
8.  Device
9.  Inlet filter