Allied International GX1660C Manual De Usuario

Descargar
Página de 41
GX1660 User Manual 
 
70-0057A-A
 
Allied Vision Technologies (Canada) Inc. 
 
 
11
Cleaning the Sensor 
 
DO NOT CONTACT CLEAN SENSOR UNLESS ABSOLUTELY NECESSARY. 
Identifying Debris 
Debris on the image sensor or optical components will appear as a darkened area or smudge on 
the image that does not move as the camera is moved.  Do not confuse this with a pixel defect 
which will appear as a distinct point. 
Locating Debris 
Before attempting to clean the image sensor, it is important to first determine that the problem is 
due to debris on the sensor window.  To do this you should be viewing a uniform image, such as 
a piece of paper, with the camera. Debris will appear as a dark spot or dark region that does not 
move as the camera is moved.  To determine that the debris is not on the camera lens, rotate the 
lens independent of the  camera.   If the spot moves as the lens moves, then the object is on the 
lens  -not  on  the  image  sensor-  and  therefore  cleaning  is  not  required.    If  the  camera  has  an  IR 
filter,  then  rotate  the  IR  filter.    If  the  object  moves  then  the  particle  is  on  the  IR  filter  not  the 
sensor.    If  this  is  the  case  remove  the  IR  filter  carefully  using  a  small  flat  head  screw  driver.   
Clean  both  sides  of  the  IR  filter  using  the  same  techniques  as  explained  below  for  the  sensor 
window. 
 
DO  NOT  TOUCH  ANY  OPTICS  WITH  FINGERS.    OIL  FROM  FINGERS  CAN 
DAMAGE FRAGILE OPTICAL COATINGS. 
 
Cleaning with Air 
If it is determined that debris is on the sensor window, then remove the camera lens, and blow 
the sensor window directly with clean compressed air.  If canned air is used, do not shake or tilt 
the  can  prior  to  blowing  the  sensor.  View  a  live  image  with  the  camera  after  blowing.    If  the 
debris  is  still  there,  repeat  this  process.    Repeat  the  process  a  number  of  times  with  increased 
intensity  until  it  is  determined  that  the  particulate  cannot  be  dislodged.  If  this  is  the  case  then 
proceed to the contact cleaning technique. 
Contact Cleaning 
Only  use  this  method  as  a  last  resort.    Use  99%  laboratory  quality  isopropyl  alcohol  and  clean 
cotton swabs.  Dampen the swab in the alcohol and gently wipe the sensor in a single stroke.  Do 
not reuse the same swab.  Do not wipe the sensor if the sensor and swab are both dry. You must 
wipe the sensor quickly after immersion in the alcohol, or glue from the swab will contaminate 
the sensor window.  Repeat this process until the debris is gone.  If this process fails to remove 
the debris, then contact AVT Canada Inc.