Battery Testing   
A DC load is well suited to verify that a battery is working properly. Only with the correct load testing, the battery can 
be confirmed if it was being the expectant life curve location. The electronic loads can be used to test any type of 
battery. 
As the accumulator is used by any sheltered equipment or the uninterrupted service system, it is necessary to use 
load testing. Because the battery is the weakest component, load testing should be used periodically to ensure the 
security of the battery. 
Capability test   
For the capability test CC mode is used.    Generate a program to set the voltage in off state. If the accumulator’s voltage 
is too low, the electronic load will interrupt the test if system checks the accumulator which is near to one rating or in 
insecurity state. In testing procedure, you could see the accumulator voltage, discharge current, electronic load power 
and spare capability of accumulator. If connecting with PC software, the battery’s discharge curve could be displayed in 
window. This test can reflect the reliability and usage time of accumulator. 
 
Operation: 
(1) Turn off the input of electronic load, then connect the load with battery. 
(2) Press
I-set
I-set
  on the panel, VFD displays "CURRENT=   A", set discharge current value and press 
Enter
to 
confirm. 
     Note: the discharge current should not be more than the current that the battery can supply. 
(3) Press 
Shift
+
Battery
, VFD displays "MIN VOLT=   V", set the min voltage value, press 
Enter
 to start 
the test.   When the battery voltage reaches as the min voltage, the battery test will stop. During the testing, press 
the down key▼and up key▲ to observe the voltage, current and discharge capacity of the battery, and the power 
of the load. 
(4)   Press 
Shift
+
Battery
to stop battery test.