SETUP Menu 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
104 
5.7. SETUP – COMM: COMMUNICATIONS PORTS 
This section introduces the communications setup menu; Section 6 provides the setup  
instructions and operation information.  Press SETUP>ENTR>MORE>COMM to arrive 
at the communications menu. 
 
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X  
 
 < TST  TST >  CAL  
 
 
SETUP
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  
RNGE
  
PASS  CLK  
MORE
  
   EXIT
 
SETUP X.X 
 
      
SECONDARY SETUP MENU
 
  
 
 
COMM
   VARS   DIAG 
 
 
   EXIT
 
 
 
 
SETUP X.X       
COMMUNICATIONS MENU 
 
  ID        INET
      
 COM1   COM2
      
                   EXIT
 
If the default password 818 
is replaced by 000, then 
Password Protection has 
been enabled. Refer to  
SETUP: PASS
 
Figure 5-13: 
SETUP – COMM Menu 
 
5.7.1. ID (INSTRUMENT IDENTIFICATION) 
Press ID to display and/or change the Machine ID, which must be changed to a unique 
identifier (number) when more than one instrument of the same model is used in a 
multidrop configuration (0) or when applying MODBUS protocol (Section 6.6.1). The 
default ID is the same as the model number; for the Model T100, the ID is 0100. Press 
any button(s) in the MACHINE ID menu (Figure 5-14) until the Machine ID Parameter 
field displays the desired identifier. 