Bacharach INSIGHT Manual De Usuario

Descargar
Página de 31
Instruction 24-9460
24
Fyrite
®
 INSIGHT
Instruction 24-9460
25
Fyrite
®
 INSIGHT
4.6 Making a Draft / Pressure Measurement
The difference in pressure (ΔP) between two areas can be measured by 
using the analyzer's two pressure ports and the PRESSURE screen.  By 
using the -ΔP port as the reference, the pressure applied to the +ΔP port 
will be displayed on the PRESSURE screen as the differential pressure 
between the two ports.  Perform a draft / pressure measurement as follows:
1. Turn ON the analyzer and allow it to complete its warm-up cycle.
2. Display the MAIN MENU by pressing the MENU (F2) button.  If 
necessary, press ESC until MENU appears 
above F2.
3. Use the ▲▼ buttons to select PRESSURE. 
Press ENTER to display the Pressure 
screen.
4. Before taking a measurement, the pressure 
sensor may need to be re-zeroed if it is not 
already displaying zero with both pressure 
ports open to the atmosphere.  If necessary, 
zero the pressure sensor as follows:
Press the ZERO (F2) button.
Disconnect any hoses connected to the 
+ΔP and -ΔP ports, and then press EN-
TER
 to zero the pressure sensor.
Reconnect any hoses.  When measuring 
draft, leave the -ΔP port open to the at-
mosphere and connect the probe's draft 
hose to the +ΔP port.
5. Do one of the following to measure draft or 
differential pressure:
To measure draft, insert the probe into 
the stack and observe the draft reading 
on the PRESSURE screen.
To measure differential pressure, con-
nect two sampling hoses to the +ΔP 
and -ΔP ports, and place the open end 
of each hose into the areas being mea-
F1        F2         F3
Atmospheric Burner 
or Gravity Vented 
Forced Air
80% Eff. Fan Assist 
or Power Vented  
Furnace/Boiler
90% Eff. Condensing 
Furnace/Boiler
Atmospheric or 
Gravity Vented 
Boiler 
Hot Water Tank
Undiluted Flue
Gas Sample Taken
Under Draft
Diverter in Top
of Fire Tube
Fig. 4.1